Tube seasoning procedure for end user

See: For Summit Synergy HF (00210,02968,L300,L500,L550,03900,03901): [url]https://summitxray.com/community/viewtopic.php?f=51&t=178&p=178&hilit=seasoning#p178[/url] For Sedecal SHFR: [url]https://summitxray.com/community/viewtopic.php?f=54&t=179&p=179&hilit=seasoning#p179[/url] For CPI CMP 200: [url]https://summitxray.com/community/viewtopic.php?f=55&t=180&p=180&hilit=seasoning#p180[/url]

Read More

What is the lifespan of an x-ray tube?

In general an x-ray tube is just an incandescent light bulb. It’s filament has a short lifetime. No amount of tube maintenance will prevent that. Filaments can withstand ~20,000 (Toshiba Tubes) to ~60,000 (Varian Tubes) exposure cycles (with a +/-10,000 variance). Data above based on internal testing performed at Summit Industries on Toshiba 7239 tubes…

Read More

What causes an open filament?

An open filament can be caused by a number of things (from most likely to least likely): Improper connection between the HV cable and the cathode receptacle. An uncalibrated generator (leading edge mA filament starting values). A worn out filament (either a very old tube or a high capacity clinic). A tube not properly maintained…

Read More

How often should a tube be seasoned?

A tube should be seasoned weekly. The #1 reason for generator failure is due to a tube that was not maintained by the end user. At some point the end user will select a technique that is abnormal for the clinic, this will result in a tube arc. The tube arc may cause damage to…

Read More

How to properly attach a HV cable to the X-ray tube

The following are excerpts from the 08538 Quick Start Guide: X-ray Tube Installation section. Apply HV Silicone grease: Tighten ring nut using hand and 1.5mm hex key per image and video below. Stop when the ring nut flange is 3mm from the HV receptacle: How to tighten HV cable ring nut:

Read More