Sedecal SHFR Error Codes
Error Codes, Generator, Sedecal SHFR:
E001
BACKUP TIMER — I2C ERROR
I2C bus error while trying to access the external redundant backup timer.
E002
WRONG WS CONFIGURATION ERROR
One or more workstations are not properly configured; a default value has been assigned.
E003
NO WS CONFIGURED ERROR
No tube has been configured for any of the workstations, there is no workstation available and a default value has been assigned.
E004
FLUORO ORDER ERROR
The Fluoro order input signal is active during the startup sequence.
E005
EXP ORDER ERROR
The Exposure order input signal is active during the startup sequence.
E006
PREP ORDER ERROR
The Preparation order input signal is active during the startup sequence.
E007
TUBE2 MODEL ERROR
The tube index (related to a tube in the tube list) configured for the tube 2 is outside boundaries, a default value has been assigned.
E008
TUBE1 MODEL ERROR
The tube index (related to a tube in the tube list) configured for the tube 2 is outside boundaries, a default value has been assigned.
E009
INVERTER ERROR (IGBT FAULT)
The inverter module has been overloaded.
E010
EEPROM CHECK ERROR
Erroneous data stored in the EEPROM.
E011
LOAD CAPACITOR ERROR
Error while charging the load capacitors. The DC-BUS voltage does not reach the right value during startup.
E012
mA RANGE ERROR
Tube current out of range during exposure.
E013
kVp RANGE ERROR
Anode-Cathode voltage out of range during exposure.
E014
kVp RAMP ERROR
Anode-Cathode voltage does not reach the final value in the designated rise time.
E015
LARGE FIL CURRENT RANGE ERROR
Large filament current out of range.
E016
SMALL FIL CURRENT RANGE ERROR
Small filament current out of range.
E018
ROTOR ERROR
The Anode Rotor Controller (starter) is not sending back the Ready condition within the designated time.
E019
mA WITHOUT EXP ERROR
Tube current without exposure order from the microcontroller.
E020
kVp WITHOUT EXP ERROR
Anode-Cathode voltage without exposure order from the microcontroller.
E023
EEPROM ERROR
Error while writing in the EEPROM.
E024
BUCKY/DIGITAL PANEL ERROR
The designated time has elapsed since the exposure signal was activated while the equipment was “Ready” without any acknowledge from the WorkStation configured as Bucky or Digital Panel.
E025
LARGE FIL DEMAND ERROR
Large filament current demand above the limit.
E026
SMALL FIL DEMAND ERROR
Small filament current demand above the limit.
E027
kVp POTENTIOMETER — I2C ERROR
I2C bus error while trying to access the digital potentiometer that adjusts the kV oscillator.
E028
ABC POTENTIOMETER — I2C ERROR
I2C bus error while trying to access the digital potentiometers that adjust the ABC window.
E029
GENERATOR OVERHEAT ERROR
Generator heat capacity exceeded. “Warning”.
E030
CORRUPT RTC ERROR
Wrong date stored in the Real Time Clock (RTC) and/or in the time stamp.
E031
TIME STAMP CHECK ERROR
The time stamp checksum is wrong.
E032
RTC—I2C ERROR
I2C bus error while trying to access the Real Time Clock (RTC).
E033
COMMUNICATIONS ERROR
The remote console has lost communication with the generator.
E034
TANK PRESOSTAT ERROR
Tank presostat opened.
E035
BUCKY MOTION (X-RAY ACK) ERROR
The acknowledge for X-rays from the Bucky or FPD has been lost before the end of the exposure.
E036
TUBE THERMOSTAT ERROR
Tube thermostat opened. “Warning”.
E037
TUBE OVERLOAD ERROR
Tube ratings exceeded or not enough Heat Units to perform the selected exposure. “Warning”.
E038
+5V POWER SUPPLY FAILURE
+5V power supply is out of range.
E039
+15V POWER SUPPLY FAILURE
+15V power supply is out of range.
E040
IMBALANCED kVp ERROR
Imbalanced kVp, there is not the same voltage in Anode and Cathode branches.
E041
IMBALANCED mA ERROR
Imbalanced mA, there is not the same current in Anode and Cathode branches.
E042
CORRUPT COUNTERS ERROR
The counters checksum is wrong.
E043
CORRUPT ERROR LOG ERROR
The error log checksum is wrong.
E044
EEPROM — I2C ERROR
I2C bus error while trying to access the EEPROM.
E045
CORRUPT TUBE DATA ERROR
The tube data checksum is wrong.
E046
BUSY BUS — I2C ERROR
I2C bus error, the bus remains always busy.
E047
LICENSE — I2C ERROR
I2C bus error while trying to access the Licence.
E048
DOOR ABORTED EXPOSURE
The door switch has been opened before the end of the exposure.
E049
COMMUNICATIONS ERROR
The generator has lost the communications with the remote console.
E050
ABORTED EXPOSURE ERROR
The user has released the exposure device/s before the end of the exposure (Prep and/or Exp).
E051
WRONG EXPOSURE TIME ERROR
The exposure time is above the maximum (very unlikely) or it is so short that it is not reachable according to the present configuration (rise time and kVp decay/high voltage capacity).
E053
FLUORO SYNC ERROR
The timeout for receiving the fluoro synchronism pulse has elapsed.
E054
DIGITAL SYNC ERROR
The timeout for receiving the Digital/DSI synchronism pulse has elapsed.
E055
NOT ENOUGH DOSE ERROR
The backup timer has elapsed before the AEC or the System has ended the exposure.
E056
NOT ENOUGH BACKUP TIME ERROR
The backup timer has elapsed before the Tomograph ends the exposure.
E057
DUAL ENERGY ERROR
It is not possible to load the Dual Energy parameters for the next exposure.
E058
TUBE1 DATA ERROR
The tube data pointed by the tube 1 index are not defined, a default tube has been selected.
E059
TUBE2 DATA ERROR
The tube data pointed by the tube 2 index are not defined, a default tube has been selected.
E060
AUTOCAL ERROR
The number of exposures to autocalibrate a mA station has run out.
E061
LICENCE ERROR
There has been an error while trying to access the Licence data. Default options have been selected.
E062
AEC ERROR
AEC selection error.
E063
ROTOR READY ERROR
The Ready from the starter has been lost before the end of the exposure.
E064
TANK FEEDBACK ERROR
The Feedback connector from the tank is not plugged in.
E065
+24V DELAYED POWER SUPPLY FAILURE
+24V delayed power supply is out of range.
E066
+24V (UNR) POWER SUPPLY FAILURE
+24V (unregulated) power supply is out of range.
E067
–15V POWER SUPPLY FAILURE
–15V power supply is out of range.
E068
+3.3V POWER SUPPLY FAILURE
+3.3V power supply is out of range.
E069
+24 PERMANENT (UNR) POWER SUPPLY FAILURE
+24 permanent (unregulated) power supply is out of range.
E070
AEC RAPID TERMINATION
Exposure was aborted because a lack of radiation received on the AEC.
E071
INTERLOCK ERROR
Exposure was aborted because in interlock configured to abort exposures has been deactivated during the
exposure.
E072
POSITIONER OK ERROR
Exposure was aborted because the signal Positioner OK (configured to be used as an interlock) has been
deactivated during the exposure.
E073
XON FEEDBACK ERROR
Exposure was aborted because the signal XON Feedback has not been activated or has been deactivated
during the exposure.
E074
COP GENERATOR RESET ERROR
The generator has been reset because of the COP module.
E075
CLK GENERATOR RESET ERROR
The generator has been reset because of the CLK module.
E076
TRAP GENERATOR RESET ERROR
The generator has been reset because of an illegal operation code.
E077
SOFTWARE INTERRUPT GENERATOR RESET ERROR
The generator has been reset because of a software interrupt.
E078
MEMORY OVERFLOW INTERRUPT GENERATOR RESET ERROR
The generator has been reset because of a memory overflow interrupt.
E079
REQUIRED MA STATIONS CAL ERROR
The required mA stations have not been calibrated.
E091
HEARTBEAT ERROR (R2CP Protocol)
The main controller is not receiving the Heartbeat messages within the configured timeout. R2CP (CAN) protocol functionality.
E099
INCORRECT MESSAGE
Last message was not recognized as a SHFR protocol message.
E101
NO SPEED AVAILABLE
The starter does not allow starting the tube neither in high speed nor in low one.
E102
LOW SPEED UNAVAILABLE
The starter does not allow starting the tube in low speed.
E103
HIGH SPEED UNAVAILABLE
The starter does not allow starting the tube in high speed.
E104
DOSIMETER TUBE 1 NO ANSWER ERROR
The dosimeter of the tube 1 doesn’t respond.
E105
DOSIMETER TUBE 1 TEST ERROR
The dosimeter of the tube 1 didn’t complete initialization test.
E106
DOSIMETER TUBE 1 STATUS ERROR
The dosimeter of the tube 1 answered a wrong or unexpected status.
E107
DOSIMETER TUBE 2 NO ANSWER ERROR
The dosimeter of the tube 2 doesn’t respond.
E108
DOSIMETER TUBE 2 TEST ERROR
The dosimeter of the tube 2 didn’t complete initialization test.
E109
DOSIMETER TUBE 2 STATUS ERROR
The dosimeter of the tube 2 answered a wrong or unexpected status.
E125 to E240 System failure related to the S0008009 Dual Speed Starter.