Synergy HF Error Codes

Error Codes, Generator, Synergy HF (00210,02968, L300, L500, L550, 03900, 03901)

CODES:
E01 kVp LIMIT
Operator selected value out of range
E02 mAs LIMIT
Operator selected value out of range
E03 TUBE LIMIT
Operator selected value out of range
E04 kW LIMIT
Operator selected value out of range
E05 TIME LIMIT
Operator selected value > 5 seconds
E06 TIME LIMIT
Operator selected value < 0.004 second E07 BUTTON Press another button to make desired change E08 mA LIMIT User attempted to select mA that is out of range for the system E09 cm LIMIT User attempted to select a cm value beyond AP data E10 MICRO-P Electrical noise, HV arcing, +5v supply, System Control PCBA E11 NO COMM. Communication cables E12 +/-12V LIMIT Power supply pcba or DC fusing E13 EEP R/W The calibration data (checked during operation) is corrupted E14 EEP CHECKSUM Calibration data (checked at power up) is corrupted E15 PREP PRESSED Prep signal active stuck ON at power up E16 EXP PRESSED Expose signal active stuck ON at power up E17 SYS COOL DOWN Wait 10 times exposure length before next exp E18 APR DATA LMT Receptor configuration at odds with AP receptor selection E19 DENSITY LMT User attempted to select >75% change in density
E20 DOOR INTLK
Open connection or opto chip on System Control PCBA
E21 TUBE INTLK
Open connection or opto chip on System Control PCBA
E22 COLMTR INTLK
Open connection or opto chip on System Control PCBA
E23 GENRL INTLK
Open connection or opto chip on System Control PCBA
E24 FAIL KV FB
Anode/Cathode kV feedback missing or wrong polarity
E25 DC BUS LOW
Checked at prep; not enough capacitor bank voltage.
E26 STILL ANODE
No rotor current during prep
E27 ANODE ROT.
Rotor current sensed when not in prep
E28 PREP HELD
Prep switch or operator
E29 NO BUCKY MOT
No B1/B2 closure
E30 mA PRESENT
Check System Control PCBA at H9 pins 5 and 6
E31 kVp PRESENT
Check System Control PCBA at H9 pins 1 and 3
E32 LOW FIL AMPS
Open secondary, Filament pcba, or poor mA leading edge calibration
E33 HI FIL AMPS
Filament pcba, or poor mA leading edge calibration
E34 IPM OVERLOAD
Excessive current through inverter
E35 EXP. RELEASE
Prep/exp switch or operator error
E36 LOW AEC RAMP
Ion chamber signal too low
E37 kVp OVERLOAD
Secondary arcing, kVp > 135
E38 mA OVERLOAD
Secondary arcing, calibration, mA is > 380 (for 300mA HF Gen.) or >570 (for 500mA HF Gen.)
E39 BACKUP MAS
Back up mAs too low
E40 NO WALL AEC
Receptor configuration at odds with AP receptor selection
E41 NO TABLE AEC
Receptor configuration at odds with AP receptor selection
E42 NO T-TOP AEC
Receptor configuration at odds with AP receptor selection
E43 SOFTWARE NOT COMPATIBLE
Controller software not compatible with AP console software
E44 KVP TOO LOW
Fail secondary or IGBTs
E45 NO ZCO
Missing ZCO signals
E46 KEY CODE BRD
Wrong Key Code board used
E47 X-RAY TUBE
Wrong Tube (or no tube) selected
E48 kV LIMIT
Selected kV not calibrated
E49 CONSOLE RST! PLS. RELEASE PREP/EXP SW!
Console power was recycled during prep
E50 CPU ERROR! WAIT FOR RST FROM CPU
Digital workstation error
E51 NO MEM CARD
User error in database transfer operation
E52 CAL DATA LMT
Attempted data entry beyond limits
E53 APR CHECKSUM
A change in AP data since last power up
E54 GEN MOD. NO.
EEPROM data corrupted
E55 AUTO CAL FAILED
Generator malfunction during Auto CAL
E56 AUTO CAL ABORTED! Or TUBE SEASON ABORTED!
Auto CAL or Tube Season stopped prematurely by user
E57 EXPOSE HELD
Expose held down for > 10 seconds after start of Auto CAL or Tube Season
E58 mA TOO LOW
Anode and Cathode cables reversed
E59 FIL I SENSOR
Filament pcba
E60 CPU COMM. ATTACK
Digital workstation error
E61 mA LOW PRE-CAL (LG)
Filament pcba, HV feedback cable or tube
E62 mA LOW PRE-CAL (SM)
Filament pcba, HV feedback cable or tube
E63 mA CAL LARGE FOCUS
Filament pcba, HV feedback cable or tube
E64 mA CAL SMALL FOCUS
Filament pcba, HV feedback cable or tube
E65 mA LOW T-SEASON CHK
Filament pcba, HV feedback cable or tube
E66 mA LOW T-SEASONING
Filament pcba, HV feedback cable or tube
E67 mA LOW SYS PWR TEST
Filament pcba, HV feedback cable or tube
E68 AC LINE VOLT NOT SELECTED
No AC Line Voltage selected
E69 K904 VER. NOT SELECTED
No K904 Version selected
E70 ERR. AC LINE OR K904 VER.
AC line voltage conflicts with K904 version
E71 mA FEEDBACK DISABLED
mA feedback turned off
E72 mAs LIMIT FOR 208 VAC
Selected mA and Exp. Time over mAs Limit for 208 VAC line
E73 MAX mA LIMIT EXCEEDED
Selected mA over Max. mA Limit
E74 mA NOT CALIBRATED
Selected mA not calibrated